(2)Eddy thickness measurement method can measure the thickness of non - conductive coatings on non - magnetic metallic substrates( such as aluminum, copper, stainless steel, etc)
Product Specificaton:
F prone
N prone
Measurement principle
magnetic induction
Eddy induction
Measurement range
0-1300um(0-1.3mm)
0-1300um(0-1.3mm)
deviation
±(3%+2um)
±(3%+2um)
accuracy
0um-99um(0.1um),100um-999um(1um),1000um-1300um(0.01mm)
calibration
At most 4 points calibration;Zero operation support
Data packet
One direct group(reading not save),4 General group(reading is saved automatically)
remark:Each group has a separate calibration setting,alarm highest and lowest setting,statistics display
Statistics value
average value,minimum value,maximum value,standard variance
unit
Um,mm,mils
alarm
highest and lowest alarm values can be set,when the values exceed highest and lowest value,LCD screen will display alarm symbol
The minimum radius of convexity
1.5mm
The minimum radius of concavity
25mm
Minimum measurement area
Diameter 5mm
Minimum substrate thickness
0.5mm
0.3mm
reading rate
A reading per second
computer interface
it is supported to download data through the USB interface
power
two battery:1.5V AAA
Operation environment
Temperature 0-40℃;humidity 20%-90%
Preservation environment
-20 to 70℃
dimension
110mm*53mm*24mm
Material and weight
ABS;84g